Versatile pH/ORP electrode with refillable reference system and integrated temperature sensor

  •  Refillable reference chamber enables the use of specific electrolytes to suit specific applications 
  • Open reference design allows for up to 10 bar overpressure (up to 145 psig) 
  • Triple ceramic diaphragms resist fouling in harsh industrial conditions 
  • Durable 0-14 pH sensitive glass can withstand CIP treatment and operating temperatures up to 130 °C (266 °F) 
  • Pt 100 or Pt 1000 RTD in Model OPS41 pH electrode increases measurement accuracy at extreme temperatures 
  • ORP sensor has platinum ring electrode as standard 
  • NEMA 6 rated detachable cable system provides the convenience of a quick connect with the reliability of a fixed cable 
  • Each electrode is serialized and performance tested at the factory
Model OPS41 pHSpecifications
Measurement rangepH: 0 – 14
ORP: -1500 to + 1500 mV
Maximum measurement error± 0.05 pH (OPS41)
± 3 mV (OPS42)
Pressure range≤ 10 bar (≤ 145 psig) with backpressure via separate KCI vessel
Temperature range-10 to 135 OC (14 to 275 OF)
Minimum conductivity10 μS/cm
Electrolyte3 mol KCl liquid, AgCI-free
KCL consumption (@ 3 psig overpressure)4.5 ml/day
Response time (@ 25 °C / 77 °F)98 % of reading in 30 seconds


Warning: call_user_func() expects parameter 1 to be a valid callback, no array or string given in /home/customer/www/analysis-oem.com/public_html/wp-content/themes/barberry/woocommerce/single-product/tabs/tabs.php on line 62

Warning: call_user_func() expects parameter 1 to be a valid callback, no array or string given in /home/customer/www/analysis-oem.com/public_html/wp-content/themes/barberry/woocommerce/single-product/tabs/tabs.php on line 62
SKU: 94b3a482d990-1-2 Categories: ,
  • Specifications
  • Configurator
Model OPS41 pHSpecifications
Measurement rangepH: 0 – 14
ORP: -1500 to + 1500 mV
Maximum measurement error± 0.05 pH (OPS41)
± 3 mV (OPS42)
Pressure range≤ 10 bar (≤ 145 psig) with backpressure via separate KCI vessel
Temperature range-10 to 135 OC (14 to 275 OF)
Minimum conductivity10 μS/cm
Electrolyte3 mol KCl liquid, AgCI-free
KCL consumption (@ 3 psig overpressure)4.5 ml/day
Response time (@ 25 °C / 77 °F)98 % of reading in 30 seconds